iStrep
A Twist Kits - Waived
Sensitivity 90%
Specificity 94%
| Each
Kit Contains: |
| |
 |
25
Test Strips |
| |
 |
25
Extraction Tubes |
| |
 |
1
Procedure Card |
| |
 |
1
Reagent A |
| |
 |
1
Reagent B |
| |
 |
1
Positive Control |
| |
 |
1
Negative Control |
|
Twist
Kit Testing Prodedure: Remove test
device from the foil pouch. Place test device on a clean, dry, level
surface. Make sure that the device is in the closed valve position,
as shown in image below.

1.
- Holding
bottles upright, add 5 drops of Reagent A.
- Then
add 5 Drops of Reagent B to the chamber.
2. 
- Add
swab to chamber
- Agitate
swab vigorously in the solution 10 times
- Let
stand 1 minute
3.
To
Remove Swab:
- Hold
down chamber with thumb and index finger.
- Lift
swab halfway up the chamber and press against the ribs inside
the wall of the chamber.
- Rotate
swab 5 times while pressing firmly against ribs to extract liquid.
-
Discard swab.
4.
- Hold
down device on a flat surface. Open Valve by twisting ribbed part
of Valve clockwise until it stops (Open Valve Position).
-
Begin timer.
- Read
results at 5 minutes.
Result
Interpretation

|